Description
The DC 10000 consists of four cascaded special dilution systems with the dilution factor 1:10. The DC 10000 dilution cascade has one aerosol inlet and four aerosol outlets. Depending on which aerosol outlet is connected to the measuring device, the aerosol with the dilution factors 1:10, 1:100, 1:1,000 or 1:10,000 are diluted and measured.
The DC 10000 can be operated with all common optical particle counters (OPC) according to ISO 12501-4 or optical aerosol spectrometers (OAS) according to ISO 12501-1. The DC 10000 can be used up to a particle size of approx. 5 µm.
Fig. 1: Overview of the DC 10000, front view
Functional principle
Fig. 2: Functional principle
Particle-free air with the volume flow VR circulates through an annular passage around the suction nozzle. Thus, according to Bernoulli, a volume flow VAn is generated in the suction nozzle.
The dilution factor VFÂ is calculated according to the formula:
Fig. 3: Formula for the dilution factor VF.
The DC 10000 needs no compressed air connection. So only an electrical power supply is necessary for operation.
Table 1: Technical characteristics of Palas® dilution systems