Test of respiratory masks better than the standard. Exact analysis of filter mask efficiency from 100 nm up to 40 µm.  SARS-CoV-2 size approx. 120 nm – 160 nm.


•      Test rig working principle better than GB 2626, EN 143, EN 149 and EN 13274-7
•      Testing of fractional efficiency, e.g. efficiency in whole size range of 100 nm up to 40 µm
•      Exact analysis of filter and filter mask efficiency for Corona Virus (size approx. 120 nm up to 160 nm) in the size range between 100nm and 180 nm we have 8 size channels
•      Future proof: Works with any kind of aerosol without adjustments
•      Further measurement of differential pressure, e.g. as well within different face velocities
•      Face velocity adjustable between 1.5 - 50 cm/s
•      Product capable of fast quality assurance AND continuous optimization in R&D (display of size distribution)
•      Individual face mask adapter for your product
•      Attractive 2 years maintenance package for availability of test rig

Description

Test of respiratory masks better than the standard with additionally exact analysis of filter mask efficiency for SARS-CoV-2 (size approx. 120 nm up to 160 nm). 8 size channels for efficiency from 100nm and 180 nm.

•  Test rig working principle better than EN 143, EN 149 and EN 13274-7
•  Equivalent to GB 2626 and 42 CFR 84 by additional software option
•  Fullfills additionally the requirements according to GB 2626 (Filter efficiency, CD 2000 discharge unit is required)
•  Testing of fractional efficiency, e.g. efficiency in whole size range of 100 nm up to 40 µm
•  Exact analysis of filter and filter mask efficiency for SARS-CoV-2 (size approx. 120 nm up to 160 nm) in the size range between 100nm and 180 nm we have 8 size channels
•  Future proof: Works with any kind of aerosol without adjustments
•  Further measurement of differential pressure, e.g. as well within different face velocities
•  Face velocity adjustable between 1.5 – 50 cm/s
Product capable of fast quality assurance AND continuous optimization in R&D (display of size distribution)
•  Individual face mask adapter for your product
•  Attractive 2 years maintenance package for availability of test rig

PMFT 1000 is based on Palas® MFP 1000.

Pictured: Analysis of filter and filter mask efficiency for Corona Virus

The size distribution of the test aerosol according to the standard is as follows:

Oil: Geom. standard deviation 1,8 | Median diameter 301 nm

Pictured: DEHS size distribution

NaCl: Geom. standard deviation 2,1 | Median diameter 63nm

Pictured: NaCl size distribution

Overview of standards for face mask penetration testing

EN 149
EN 13274-7
EN 13274-7
GB 2626
GB 2626
42 CFR 84
42 CFR 84
Aerosol
see EN 13274-7
NaCl
PaO
NaCl
PaO/DOP
NaCl
DOP
Mean diameter
see EN 13274-7
0.06 – 0.1 µm
0.29 – 0.45 µm
0.055 – 0.095 µm
0.165 – 0.205 µm
0.055 – 0.095 µm
0.165 – 0.205 µm
Standard deviation
see EN 13274-7
2 – 3
1.6 – 2.2
< 1.86 (by additional software module)
< 1.6 (by additional software module)
< 1.86 (by additional software module)
< 1.6 (by additional software module)
Concentration
see EN 13274-7
4 –12 mg/m³
15 – 25 mg/m³
< 200 mg/m³
(50 mg/m³) < 200 mg/m³
< 200 mg/m³
< 200 mg/m³
Discharge
required
required
required
required
Air flow
see EN 13274-7
95 l/min
95 l/min
85 ± 4 l/min
85 ± 4 l/min
85 ± 4 l/min
85 ± 4.25 l/min
Temperature
see EN 13274-7
22  ± 3 °C
25 ± 5 °C
25 ± 5 °C
25 ± 5 °C
25 ± 5 °C
Rel. humidity
see EN 13274-7
< 40 %

20 – 40 %

(by compressed air)

20 – 40 %

(by compressed air)

20 – 40 %

(by compressed air)

Measurement device
see EN 13274-7

Sodium flame photometer

Light scattering photometer

particle detector

particle detector

Light scattering photometer

Light scattering photometer

Measuring time
see EN 13274-7
30 s
30 s

lowest eff. during loading

lowest eff. during loading

lowest eff. during loading

lowest eff. during loading

Pause time
see EN 13274-7
180 s
180 s

lowest eff. during loading

lowest eff. during loading

lowest eff. during loading

lowest eff. during loading

Exposition
120 mg
120 mg
120 mg

200 ± 5 mg

200 ± 5 mg

200 ± 5 mg

200 ± 5 mg

PMFT remarks

O.K.

O.K.

O.K.

O.K. with upgrade KIT

O.K. with upgrade KIT

O.K. with upgrade KIT

O.K. with upgrade KIT

Datasheet

Measurement range (size)

0,10 – 40 µm

Volume flow

1 – 27 m3/h (Druckbetrieb)

Power supply

115/230 V, 50/60 Hz

Dimensions

approx. 600 • 1,800 • 900 mm (W • H • D)

Installation conditions

19 –23 °C

Inflow velocity

5 – 100 cm/s (others on request)

Differential pressure measurement

0 – 1200 Pa

Test area of the medium

100 cm²

Aerosols

Dusts (e. g. SAE dusts), salts (e. g. NaCl, KCl), liquid aerosols (e. g. DEHS)

Aerosol concentrations

For SAE Fine without additional dilution up to 1,000 mg/m3 (ISO A2 Fine)

Compressed air supply

6 – 8 bar

Benefits


•      Test rig working principle better than GB 2626, EN 143, EN 149 and EN 13274-7
•      Testing of fractional efficiency, e.g. efficiency in whole size range of 100 nm up to 40 µm
•      Exact analysis of filter and filter mask efficiency for Corona Virus (size approx. 120 nm up to 160 nm) in the size range between 100nm and 180 nm we have 8 size channels
•      Future proof: Works with any kind of aerosol without adjustments
•      Further measurement of differential pressure, e.g. as well within different face velocities
•      Face velocity adjustable between 1.5 - 50 cm/s
•      Product capable of fast quality assurance AND continuous optimization in R&D (display of size distribution)
•      Individual face mask adapter for your product
•      Attractive 2 years maintenance package for availability of test rig

Applications


•      Test of respiratory masks
•      Exact analysis of filter mask efficiency for e.g. Corona Virus
•      Filter testing for HEPA quality