Separate dispersion and control unit, pressure resistant at over-pressure values up to 3 bar, optional low pressure operation from 300 mbar (absolute pressure), nitrogen as dispersion gas as well.

•    Pressure-resistant up to 3 barg overpressure
•    2 m distance between dispersing unit and control unit
•    Optional:Low pressure operation from 300 mbar absolute
•    Nitrogen as dispersing gas
•    Optional: Remote control or computer-controlled

Description

This device disperses particles at positive pressure values of up to 3 bar and can also use nitrogen, in addition to air, as the dispersing gas.

Optional operation with low pressure from 300 mbar absolute is possible (please see accessories).

Please note: The 7-, 10-, 14- or 20-mm feed stock reservoirs are pressure-resistant.

For operation with low pressure special pressure-resistant feed stock reservoirs are needed. Their piston is strongly connected to the feeding unit by a claw. This enables an undisturbed operation with low pressure. Old RBG models can be upgraded with this function by Palas®.

The solid material reservoir with a diameter of 28 mm is not pressure-resistant, but can be used with the RBG 1000 SD under atmospheric conditions.

K01985
Table 1: Mass flows of RBG system (compacted density 1 g/cm3)
0r1982
Table 2: Dispersion covers
DT1984
Table 3: Different versions of the RBG system

I = version for inhalation
D = pressure-resistant
G = low feed rate
L = easily removable and weighable dosing unit
S = nitrogen version

Datasheet

Volume flow

0.5 – 5.0 m3/h

Power supply

115/230 V, 50 – 60 Hz

Dimensions

430 • 300 • 180 mm (L • W • H, dispersion unit)

Weight

approx. 19 kg

Particle material

Non-cohesive powders and bulks

Dosing time

Several hours nonstop

Maximum particle number concentration

ca. 107 particles/cm3

Mass flow (particles)

0.04 – 430 g/h (with an assumed compacted density of 1 g/cm^3)

Particle size range

0.1 – 100 µm

Carrier/dispersion gas

Air, nitrogen

Pre-pressure

4 – 8 bar

Feed rate

5 – 700 mm/h

Reservoir diameter

7, 10, 14, 20 mm

Maximum counter pressure

200 mbarg

Reservoir length

70 mm

dispersion cover

Type A, type B, type C, type D

Compressed air connection

Quick coupling

Aerosol outlet connection

Dispersion cover type A: Øinside= 5 mm, Øoutside = 8 mm; Dispersion cover type B: Øinside= 3.6 mm, Øoutside = 6 mm; Dispersion cover type: Øinside= 2.5 mm, Øoutside = 6 mm

Filling quantity

2.7 g (reservoir Ø = 7 mm), 5.5 g (reservoir Ø = 10 mm), 10.8 g (reservoir Ø = 14 mm), 22 g (reservoir Ø = 20 mm), 43 g (reservoir Ø = 28 mm)

Benefits

•    Pressure-resistant up to 3 barg overpressure
•    2 m distance between dispersing unit and control unit
•    Optional:Low pressure operation from 300 mbar absolute
•    Nitrogen as dispersing gas
•    Optional: Remote control or computer-controlled

Applications

•    All applications pressure-resistant up to 3 barg overpressure
•    Dispersion of radioactive substances
•    Dispersion of pharmaceutical powders
•    Filter industry:
      -   Determination of fractional separation efficiency
      -   Determination of total separation efficiency
      -   Long-term dusting
      -   Filter media and ready-made filters
      -   Dust removal filters
      -   Vacuum cleaners and vacuum cleaner filters
      -   Car interior filters
      -   Engine air filters
•    Calibration of particle measurement devices
•    Flow visualization
•    Inhalation tests
•    Tracer particles for LDA, PIV, etc.
•    Coating of surfaces