关注微信:

產品

熱線
下載

PMPD 100

Dilution system ejector for the PMP application with dilution factor 1:100

Categories: ,

Description

The PMPD 100 dilution system is a dilution system according to the ejector principle that was especially developed for the PMP application or the PMP measurement chain. The PMPD 100 achieves a dilution factor of 1:000 (see Figure 1) by means of a thermodiluter up to 200°C.


Fig. 1: PMPD 100

The PMPD dilution systems offer all the advantages of the other Palas® product series of ejector diluters, e.g. a temporally constant dilution factor.

The suitability of the PMPD 100 for the PMP measurement chain was confirmed at the METAS Institute in Switzerland ( see measurement report no. 235-10383 ).

Representative dilution of particle size distribution of the Palas® dilution systems by cascading

VDI report no. 1973 from 2007 proved metrologically that a reproducible aerosol dilution is possible with the Palas® dilution systems down to VF 100,000.

Table 1:  Technical characteristics of Palas® dilution systems

Additional information

Benefits

<span style="color:#008b57">•</span>    The dilution systems from Palas® are characterized unambiguously. This is documented with a calibration certificate for each individual device.
<br> <span style="color:#008b57">•</span>    The dilution steps for the PMPD series deliver a temporally constant, representative dilution with the factors 100 and 1000.
<br> <span style="color:#008b57">•</span>    Low compressed air consumption (e.g. just 96 l/min. for a dilution factor of 1000 with four VKL 10 systems)
<br> <span style="color:#008b57">•</span>    The dilution steps are combinable with all common particle counters.

Applications

Dilution system for PMP measurement chain

Power supply

115 – 230 V, 50 – 60 Hz

Dilution factor

1 : 100

Isokinetic suction nozzles

2 – 5 l/min

Maximum particle size

< 10 µm

Special features

Evaporation of volatile elements for exhaust emission measurements according to VPR Calibration Procedure AEA/ED 47382/Issue 5 (Volatile Particle Removal Efficiency), chemical resistant, heated to 200 °C

Thermodynamic conditions for dilution

≤ 400°C

Volume flow (clean air)

36 – 90 l/min (heated to 200 °C)

Volume flow (suction flow)

2 – 5 l/min

Compressed air supply

4 – 8 bar