关注微信:

產品

熱線
下載

MFP 3000 C

Version for testing filter media better than DIN 71460 and ISO 11155-1 road vehicle interior filters

Description

Version MFP 3000 C is especially tailored to the requirements of the DIN 71460 and ‘ISO 11155-1 Road vehicle interior filters’ measurement procedures.

Aerosol generation with RBG 1000 G (DIN 71460) and AGK 2000 (ISO TS11155-1):

The dosing air for the RBG 1000 G and AGK 2000 solid particle aerosol generators is regulated with a mass flow controller. This ensures the same operating conditions are always maintained.

DLB 2000 compressed air humidifier for the dispersion air:

Dry compressed air is normally used for aerosol generation, whereby, at the very low volume flows in filter media testing, the rel. humidity of the test volume flow can drop considerably. The DLB 2000 compressed air humidifier can condition the rel. humidity and temperature of the RBG 1000 dispersion air precisely to the required values and thus minimises the influence of rel. humidity on the dust holding capacity to be measured.

Aerosol inlet and aerosol neutralization on MFP 3000 C:

The aerosol inlet on the MFP 3000 C is equipped with a corona discharge to neutralize the test aerosol and ensures a homogeneous distribution of the test aerosol in the raw gas channel. The simple construction allows rapid replacement of the aerosol generator and the raw gas channel is easy to clean.

welas® 2100 aerosol sensors:

The welas® 2100 high concentration sensors ensure unambiguous and coincidence-free fractional separation efficiency measurement with very good count statistics at the specified concentration of 75 mg/m3 ‘ISO A2 Fine dust’. These sensors are also fitted with a special aerosol guide that minimises contamination of the internal optics.

Software:

Various differential pressure levels can be set in the filter media test sequence program for loading in accordance with DIN 71460. The clear definition of the test parameters in the pre-programmable sequence programs ensure very high level comparability of the results.

Additional information

Benefits

<span style="color:#008b57">•</span>    Virtually simultaneous particle measurement in the raw gas and clean gas
<br> <span style="color:#008b57">•</span>    Particle size measurements from 0.2 – 40 µm
<br> <span style="color:#008b57">•</span>    Measurement of Cn max= 105 particles/cm3 without dilution
<br> <span style="color:#008b57">•</span>    Internationally comparable measurement results
<br> <span style="color:#008b57">•</span>    Widespread distribution of the measurement system
<br> <span style="color:#008b57">•</span>    High reproducibility of the testing method
<br> <span style="color:#008b57">•</span>    Easy use of different test aerosols, e.g. SAE Fine and Coarse, NaCl/KCl, DEHS
<br> <span style="color:#008b57">•</span>    Highest raw gas concentrations of up to > 100 mg/m3 (ISO Fine) or > 500 mg/m3 (ISO Coarse) with measurement of the fraction separation efficiency for burden tests
<br> <span style="color:#008b57">•</span>    Flexible filter test software FTControl
<br> <span style="color:#008b57">•</span>    Sequence programs for pressure loss measurements, measurements of fraction separation efficiency and burden measurements
<br> <span style="color:#008b57">•</span>    Easy to operate, even untrained personnel can be instructed quickly in the use of the equipment
<br> <span style="color:#008b57">•</span>    Short set-up times
<br> <span style="color:#008b57">•</span>    Cleaning and calibration can be performed autonomously by the customer
<br> <span style="color:#008b57">•</span>    Easy use of the measurement technology components – even in other applications
<br> <span style="color:#008b57">•</span>    Mobile setup, easy to move on castors
<br> <span style="color:#008b57">•</span>    Reliable operation
<br> <span style="color:#008b57">•</span>    Validation of the clear function of individual components and the overall system during pre-delivery acceptance testing and upon delivery
<br> <span style="color:#008b57">•</span>    Low-maintenance
<br> <span style="color:#008b57">•</span>    The unit will reduce your operating costs

Applications

<span style="color:#008b57">•</span>    Testing of filter media and small filter elements in product development and during production monitoring.
<br> <span style="color:#008b57">•</span>    Testing option based on ISO 11155-1 / DIN 71460-1 (cabin air filters).

Measurement range (size)

0.2 – 40 µm

Volume flow

1 – 35 m3/h (suction mode)

Dimensions

680 • 2,500 • 1,550 mm (W • H • D)

Inflow velocity

5 – 100 cm/s (others on request)

Differential pressure measurement

0 – 5,000 Pa

Test area of the medium

100 cm²

Aerosols

Dusts (e. g. SAE dusts), salts (e. g. NaCl, KCl), liquid aerosols (e. g. DEHS)

Aerosol concentrations

For SAE Fine without additional dilution up to 1,000 mg/m3 (ISO A2 Fine)

Compressed air supply

6 – 8 bar