With automatic regulation of the sampling flow through the welas® aerosol sensors at an overpressure up to 10 bar

Description

Depending on the composition of the aerosol to be measured, i.e. the carrier gas component and the particle material, pressure changes in the carrier gas can significantly influence the particle size distribution, e.g. due to condensation or evaporation.

For this reason, the welas® aerosol sensors welas® 2070 P, 2100 P, 2200 P, 2300 P and welas® 2500 P are equipped with a pressure-tight cuvette to ensure isobaric and isothermal sampling into the sensor’s measurement volume.

The Promo® is usually calibrated for the operating volume flow. As the operating volume flow changes with pressure, it is advantageous for the user if automatic volume flow regulation for the sampling volume flow is provided for in the device.

In the Promo® 2000 P the pressure of the carrier gas is measured and the required operating volume flow is automatically set to 5 l/min.

Includes:

    Mass flow controller for volume flow regulation
    Absolute pressure capsule
    Filter unit

Additional information

Benefits

<span style="color:#008b57">•</span>    Measuring range of 0.2 to 100 µm (up to 4 measuring ranges selectable in one device)
<br> <span style="color:#008b57">•</span>    Up to four measuring ranges in only one device:
<br>       <span style="color:#008b57">-</span>   0,2 µm – 10 µm
<br>       <span style="color:#008b57">-</span>   0,3 µm – 17 µm
<br>       <span style="color:#008b57">-</span>   0,6 µm – 40 µm
<br>       <span style="color:#008b57">-</span>   2 µm – 100 µm (additionally for sensors 2300 and 2500)
<br> <span style="color:#008b57">•</span>    Up to 128 size channels per measuring range
<br> <span style="color:#008b57">•</span>    Concentration range of 1 particle/cm3 to 106 particles/cm3
<br> <span style="color:#008b57">•</span>    Calibration curves for different refractive indices
<br> <span style="color:#008b57">•</span>    Very high and reproducible counting efficiency rate starting at 0.2 µm
<br> <span style="color:#008b57">•</span>    Pressure-resistant up to 10 bar (optional)
<br> <span style="color:#008b57">•</span>    Optical fibre technology
<br> <span style="color:#008b57">•</span>    Simple operation with a large touch display
<br> <span style="color:#008b57">•</span>    Calibration, cleaning and lamp replacement can all be performed independently by the customer
<br> <span style="color:#008b57">•</span>    External control by RS 232 or Ethernet
<br> <span style="color:#008b57">•</span>    With analysis software PDAnalyze
<br> <span style="color:#008b57">•</span>    Optional: Software PDControl for operation as welas® digital available
<br> <span style="color:#008b57">•</span>    Low maintenance
<br> <span style="color:#008b57">•</span>    Reliable function
<br> <span style="color:#008b57">•</span>    Reduces your operating expenses

Applications

<span style="color:#008b57">•</span>    Emission monitoring of installations
<br> <span style="color:#008b57">•</span>    Control of grinding and classification processes
<br> <span style="color:#008b57">•</span>    Monitoring of production processes in the food, pharmaceuticals and chemicals industries
<br> <span style="color:#008b57">•</span>    Testing of complete filters, inertial and wet separators or electrostatic precipitators

Interfaces

USB, Ethernet, RS232/485, Wi-Fi

Measurement range (size)

0,2 µm – 10 µm,
0,3 µm – 17 µm,
0,6 µm – 40 µm,
2 µm – 100 µm

Size channels

up to 128 (64/decade)

Measuring principle

Optical light-scattering

Measurement range (number CN)

< 1 • 106 particles/cm3

Volume flow

5 l/min regulated by mass flow

Data acquisition

20 MHz processor, 256 raw data channels, digital

Light source

Xenon arc lamp 35 W

Power consumption

100 W

User interface

Touch screen, 800 • 480 pixel, 7"

Housing

Table housing, optionally with mounting brackets for rack-mounting

Dimensions

185 • 450 • 315 mm (H • W • D) (19")

Support options

Direct remote access, Palas® webserver service

Weight

approx. 8 kg (control unit), approx. 2.8 kg (sensor)

Operating system

Windows embedded

Data logger storage

4 GB Compact Flash

Software

PDControl, FTControl, PDAnalyze

Installation conditions

+5 – +40 °C (control unit)