关注微信:

產品

熱線
下載

CIF 2000

CIF 2000 test rig for standard tests on motor vehicle passenger compartment filters in accordance with DIN 71460 part 1 and ISO/TS 11155-1

Description

Palas® offers the CIF 2000 test rig with the Promo® 2000 unit for standard tests on motor vehicle passenger compartment filters in accordance with DIN 71460 part 1 and ISO/TS 11155-1.

The volume flow is automatically regulated and adjustable from 60 to 800 m3/h.
With the aid of the CIF, both the fraction separation efficiency and the burden on the complete filters or filter media are measured and evaluated.

Particle dosing for the filter test is performed with the brush generator RBG 1000, which has been internationally established for over 30 years. On the CIF 2000, overall and fraction separation efficiency are determined with the high-resolution light scattering spectrometer Promo® 2000.

The main advantage of the Promo® 2000 systems lies in the use of fiber optic technology. Thanks to this, the sensor heads can be positioned directly at the measurement site, as a result of which line losses are minimized. With the Promo® 2000 measuring systems, reliable tests to determine the separation efficiency and burden tests can be performed across the entire measurement range for all particle sizes.

Following ISO/TS 11155-1, the test rig is also equipped with the salt aerosol generator AGK 2000. In the testing conduit, filter media up to a size of approx. 220 x 500 mm can also be tested under the conditions stated above. In order to test the influence of real ambient conditions on the separation behavior of filters, Palas®offers an upgrade to the CIF test rig with climate technology components foradjustment of the temperature between +18°C and +90°C and the relative humidity between 30% and 70%.

As an option, the test rig can be upgraded after provision of the measuring equipment for gas analysis for investigations of the adsorption and desorption dynamics following DIN 71460 part 2 and ISO/TS 11155 part 2.

Automation

The CIF is equipped with a high-pressure ventilator, which can be throttled on the pressure side with infinitely variable adjustment and is frequency-controlled, and which can be automatically regulated via the FTControl filter test software for regulation of the volume flow. Furthermore, sensor data such as the volume flow, temperature, relative humidity and differential pressure at the filter are also recorded automatically during the filter test. The aerosol generators and the corona discharge can be actuated via the FTControl software.

Measurement of the fraction separation efficiency

Fig. 1: Example showing a comparison of fraction separation efficiencies

   Clear demonstration of the separation efficiency of your filter or filter medium throughout the entire measurement range from 0.2 to 40 µm with the Promo® 2000 system
   Highest measurement reproducibility and repeatability highlight even fine differences in the separation efficiency
   Short measurement times of around 2 minutes per separation efficiency measurement thanks to optimized application
   Simple comparison of separation efficiency curves, calculation of mean values also possible

Burden / hold time measurement and record of the pressure loss curve

Fig 2: Example of the hold time measurement

   Performance of measurements of the fraction separation efficiency during dust application; pressure loss or measurement time can be pre-selected as the abort criterion
   Determination and representation of the pressure loss curve and retention curve in diagram and table form The representation of the particle diameters at 80% and 90% separation provides additional information.
   Comparison of the fraction separation efficiencies during the different loading steps
   Shortening of the measurement times, e.g. through increased aerosol concentration

Further component:

   Measurement point switching device

Additional information

Benefits

<br> <span style="color:#008b57">•</span>    Particle measurements from 200 nm
<br> <span style="color:#008b57">•</span>    Measurement and evaluation of fraction separation efficiency and burden
<br> <span style="color:#008b57">•</span>    Automatic data acquisition for barometer pressure, temperature, humidity, differential pressure
<br> <span style="color:#008b57">•</span>    Temperature control (+18°C to +90°C) and moisture control (30 – 70%)
<br> <span style="color:#008b57">•</span>    Automatic actuation of all test rig components
<br> <span style="color:#008b57">•</span>    Automatic performance of the measurement processes
<br> <span style="color:#008b57">•</span>    Individual programming of measurement processes for filter testing using the FTControl software
<br> <span style="color:#008b57">•</span>    Separate measurement and analysis parts – this saves time and money, as the analysis can be performed while the measurement is still on-going
<br> <span style="color:#008b57">•</span>    Printouts and saving of complete test records
<br> <span style="color:#008b57">•</span>    Easy access to all data for the recorded measurement signals from the up to 6 external sensors
<br> <span style="color:#008b57">•</span>    Low-maintenance
<br> <span style="color:#008b57">•</span>    Easy operation
<br> <span style="color:#008b57">•</span>    Reliable operation
<br> <span style="color:#008b57">•</span>    The unit will reduce your operating costs

Applications

<br> <span style="color:#008b57">•</span>    Complete filter test in accordance with DIN 71460 part 1 & ISO/TS 11155 part 1
<br> <span style="color:#008b57">•</span>    Test of filter media in accordance with DIN 71460 part 1 & ISO/TS 11155 part 1
<br> <span style="color:#008b57">•</span>    Testing of other complete filters and filter media

Measurement range (size)

0.2 – 40 µm

Volume flow

60 m3/h – 800 m3/h (cycle operation)

Dimensions

Test rig: 1,000 • 2,800 • 4,200 mm, filter holder: 300 • 600 mm (filter and others on request)

Material

Stainless steel V2A, 2 mm

Temperature regulation

+18°C – +90°C

Luftfeuchteregelung

30% up to 70%

Temperature- und humidity sensor

– measuring range: -20°C – +80°C, 0 – 100 % rH, – accuracy: ± 0.1°C (20°C), ± 1 % rH (0 – 90 % rH), ± 2 % rH (90 – 100 % rH)

Barometric pressure gauge

– measuring range: 600 – 1100 hPa, – accuracy: ± 0.10 hPa

Differential pressure gauge

– Messbereich: < 2500 Pa, – Linearitätsfehler: < 0,2 % vom Endwert

Measurement of the air velocity

– measuring range: 0.5 – 40 m/s, – accuracy: < +/- 0.05 m/s (up to 20 m/s), < +/- 0.08 m/s (20 – 30 m/s), < +/- 0.1 m/s (30 – 40 m/s)

Preconditions

3 phases, 400 V, neutral, earth connection of approx. 3 KW and provision of pressurized air max. 8 bar