PLG 1000

Aerosol generator for the defined atomization of oils and use on filter test rigs

Category:

Description

The PLG 1000 is a cold atomizer intended for use in air-conditioned rooms. If the room is unable to be air-conditioned, then a heatable version of the device should be used, e.g. PLG 2000 H.

Fig. 1: PLG 1000

Startup

The liquid to be dispersed is filled in the reservoir. The nozzle system developed by Palas® is immersed in the liquid. This nozzle system is based on the Laskin principle and guarantees extremely precise dosing constancy with uniform particle size. The mass flow is adjusted using the volume flow through the nozzle. The volume flow is controlled by a pressure regulator and a manometer on the device.

Fig. 2: Functional principle of the PLG series

Legend

1) Compressed air

2) Special Laskin nozzle

3) Aerosol substance

4) Aerosol outlet

This nozzle system is based on the Laskin principle and guarantees extremely precise dosing constancy with uniform particle size (see Fig. 3).

Fig. 3: Highly constant particle concentration over time at a resolution of 30 s

The PLG 1000 generates mass flows of up to approx. 5 g/h max. (depending on the aerosol substance in use).

Additional information

Benefits

<span style="color:#008b57">•</span>    Excellent short-term and long-term dosing constancy
<br> <span style="color:#008b57">•</span>    Best reproducibility with respect to particle size distribution and particle concentration
<br> <span style="color:#008b57">•</span>    Large mass volume range (very low and very high)
<br> <span style="color:#008b57">•</span>    Robust design (optionally resistant against chemically aggressive liquids)
<br> <span style="color:#008b57">•</span>    Compact and light
<br> <span style="color:#008b57">•</span>    Easy to operate, proven in industrial applications
<br> <span style="color:#008b57">•</span>    Reliable function
<br> <span style="color:#008b57">•</span>    Low maintenance

Applications

<span style="color:#008b57">•</span>    Filter industry/oil separators
<br>       <span style="color:#008b57">-</span>   Determination of separation efficiency
<br>       <span style="color:#008b57">-</span>   Determination of fractional separation efficiency
<br>       <span style="color:#008b57">-</span>   Loading test
<br>       <span style="color:#008b57">-</span>   HEPA/ULPA filters
<br> <span style="color:#008b57">•</span>    Test of cooling lubricant separators
<br> <span style="color:#008b57">•</span>    Comparison of particle measurement devices
<br> <span style="color:#008b57">•</span>    Tracer particles
<br> <span style="color:#008b57">•</span>    Flow visualization

Volume flow

1 – 23 l/min

Dimensions

280 • 130 • 100 mm (H • W • D)

Weight

approx. 2 kg

Mass flow (particles)

2.5 g/h

Aerosol outlet connection

Øinside = 11 mm,

Øoutside = 14 mm

Mean particle diameter (number)

0.4 µm (DEHS)

Filling quantity

70 ml