welas® digital 2000

Aerosol spectrometer with an aerosol sensor, highest particle size resolution, measuring range from 200 nm to 100 µm with light wave conductor technology

    Measuring range of 0.2 to 100 µm (4 measuring ranges selectable in one device)
    Up to four measuring ranges in only one device:
      -   0,2 µm – 10 µm
      -   0,3 µm – 17 µm
      -   0,6 µm – 40 µm
      -   2 µm – 100 µm (additionally for sensors 2300 and 2500)
    Up to 128 size channels per measuring range
    Concentration range of 1 particle/cm3 to 106 particles/cm3
    Calibration curves for different refractive indices
    Very high and reproducible counting efficiency rate starting at 0.2 µm (see Graph 2)
    High temporal resolution down to 10 ms
    Optical fibre technology
    Measurement in potentially explosive environment
    Long service life of the light source of 2000 h
    Extensive PDControl and FTControl software
    Simple operation
    Calibration, cleaning and lamp replacement can all be performed independently by the customer
    Low maintenance
    Reliable function
    Reduces your operating expenses

Description

The welas® digital 2000 is a flexible, powerful and economical light-scattering spectrometer system, which determines particle concentration and size precisely and reliably.

Unique are up to four measuring ranges in only one device:
    0.2 µm – 10 µm
    0.3 µm – 17 µm
    0.6 µm – 40 µm
    2 µm – 100 µm (additionally for sensors 2300 and 2500).

welas® digital 2000 is famous for up to 128 size channels per measuring range and a concentration range from < 1 particle/cm3 to 106 particles/cm3.

The device is characterized by its optical fibre technology, too. The welas® sensor is connected via a fiber optic cable with a length up to 30 m with the welas® digital control unit.

This leads to a minimization of particle losses in long sampling lines by simply installing the sensor directly at the sampling location.

Connection via fiber-optic cable allows the welas® 2000 and 2000 P series sensors to be easily connected to the control and evaluation unit and interchanged as required.

The welas® sensors are equipped with different sized measurement volumes. This allows adaptation of the measuring device to the particle concentration present in the application, such that a high counting rate can be achieved with a short measuring time.

The aerosol sensors allow reliable measurement in the concentration range from < 1 particle/cm3 up to 106 particles/cm3.

The welas® digital is based on scattered-light analysis on a single particle. In the welas® digital, the special advantages of the well-known and internationally acclaimed welas® system are combined with new and fast digital individual signal processing and coincidence correction is enabled.

The high size classification accuracy and the high size resolution are guaranteed by the following special feature (see Graph 1):

    White light and 90° light-scattering detection
Unambiguous calibration curve
    Patented T-aperture
No border zone error
    New digital individual signal processing

Coincidence detection and correction of the individual signal making it possible to measure in higher concentrations.

The welas® digital measurement technology

welas® digital offers a new, fast 20 MHz signal processing processor, which analyses the progression of each particle signal. This makes it possible to recognise coincidental events in light scattering measurement technology at the individual signal and correct them (according to Dr. Umhauer / Prof. Dr. Sachweh).

This way it is possible to increase the maximum concentration limit up to 106 particles/cm3 (welas® 2070 sensor).

Also in low concentrations < 1 particle/cm3 with the welas® 2500 sensor, this leads to a higher measuring accuracy.

High classification accuracy, high resolution capability and a high counting efficiency are the prerequisite for unambiguous particle measurement.


Fig. 1: Resolution capability and classification accuracy

The welas® digital is characterized by its very high counting efficiency starting from 0.2 µm!


Fig. 2: Example with 2200 sensor, in relation to LAS-X II

The welas® PDControl digital software

The welas® digital is controlled via a laptop using the PDControl software. The software allows particle measurements and calibration of the measurement device. In addition, the measurements can be analyzed and compared in detail with a temporal resolution down to 10 ms.

Datasheet

Interfaces

USB

Measurement range (size)

0.2 µm – 10 µm,
0.3 µm – 17 µm,
0.6 µm – 40 µm,
2 µm – 100 µm

Size channels

up to 64/decade

Measuring principle

Optical light-scattering

Measurement range (number CN)

< 1 • 106 particles/cm3

Volume flow

5 l/min

Data acquisition

20 MHz processor, 256 raw data channels, digital

Light source

Xenon arc lamp 35 W

User interface

Laptop

Housing

Table housing, optionally with mounting brackets for rack-mounting

Dimensions

185 • 450 • 315 mm (H • W • D) (19")

Weight

approx. 18 kg (control unit), approx. 2.8 kg (sensor)

Software

PDControl, FTControl

Installation conditions

+5 – +40 °C (control unit)

Benefits

    Measuring range of 0.2 to 100 µm (4 measuring ranges selectable in one device)
    Up to four measuring ranges in only one device:
      -   0,2 µm – 10 µm
      -   0,3 µm – 17 µm
      -   0,6 µm – 40 µm
      -   2 µm – 100 µm (additionally for sensors 2300 and 2500)
    Up to 128 size channels per measuring range
    Concentration range of 1 particle/cm3 to 106 particles/cm3
    Calibration curves for different refractive indices
    Very high and reproducible counting efficiency rate starting at 0.2 µm (see Graph 2)
    High temporal resolution down to 10 ms
    Optical fibre technology
    Measurement in potentially explosive environment
    Long service life of the light source of 2000 h
    Extensive PDControl and FTControl software
    Simple operation
    Calibration, cleaning and lamp replacement can all be performed independently by the customer
    Low maintenance
    Reliable function
    Reduces your operating expenses

Applications

    Determination of the separation efficiency of car interior filters, engine air filters, room air filters, compressed air filters, vacuum cleaner filters, cleanable filters, electrostatic precipitators, oil separators, cooling lubricant separators, wet scrubbers, cyclones and other separators
    Isothermal and isobaric particle size and quantitative determination, for instance in the automobile, chemical, pharmaceutical and food industries
    Analysis of fast, transient processes
    Inspection of smoke detectors
    Particle formation for cloud formation
    Emission measurements
    Immission measurements
    Breathing function: Inhalation / Exhalation (Particle size and number)